Fastcomtec_parameter_notes.md 6.45 KiB

Basic Modes of Operation

1. Stop after sweep mode (p.35f)

When the MCS8A is armed it waits for a START input signal. When one occurs the sweep is started / triggered, meaning the time starts to count. Now the arrival times of the STOP input signals relative to the start are acquired on all channels that are enabled. A STOP event can be either a falling or a rising edge or both. Even further signals into the START input can be acquired. Since the type of edge is detected and marked in the acquired data even Time-over-Threshold or pulse width measurements can be accomplished.

The data acquisition can be ended after a given time via the Time preset. In the edit field Range the length of the spectrum can be entered. A Bin width of 1 means the highest time resolution. The Bin width can be chosen in powers of 2 up to 1677216 times the elementary dwell time. If an Acq. Delay is specified, data are acquired in a sweep not before the specified time. Hold after sweep allows to wait a specified time after a sweep before the next sweep can be started.

1.1. Start of measurement

1.1.1. Endless mode (p.44)

mode of the measurement can be endless if the corresponding checkbox is crossed (in settings)

1.1.2. Software start (p.44)

Doesn't require a start signal (checkbox in settings dialogue)

1.2. Termination of measurement

terminated by any of the preset (2) conditions

1.2.1. Time preset (total acquisition time) - THIS

in steps of 2.56 ns - total measurement time (in the document: "sweep and so the data acquisition ends after that time" but with sweep preset you can get multiple sweeps. So this is more like a total measurement time)

1.2.2. Sweep preset (number of sweeps)- THIS

  • 48-bit sweep counter counts the real start of a new sweep rather than the completion of the sweeps
  • preset enabled: after preselected number of sweeps any further start of a sweep is prohibited

1.2.3. ROI preset (number of counts) - THIS

Stop measurement after certain number of counts. ROI Preset

1.3. Single sweep settings

1.3.1. Acquisiton delay (counting delay after START) - THIS

in steps of 2.56 ns - accept only STOP events (= signals) that arrive after the selected delay